Miele G 7836 Dishwasher User Manual


 
39
RO-190 M2 reverse osmosis system
• For a continuous supply of demineralised
water
• Throughput: max. permeate capacity
190 l/h
• Stand-alone solutions for installation in
neighbouring unit
• Stainless-steel panelling
• External dimensions: H 380, W 543,
D 302 mm
• Further features and technical data as per
RO-160 M2
Optional for RO-190 M2 and RO-190 M1
(VEOLIA accessory list)
• Installation of a pre-filter
• Additional connection, e.g. for connection
of autoclave or water dispenser tap
• Pressure tank to store demineralised
water
• Connection to water softener
Indicator lights
RO-190 M2 reverse osmosis system
• For a continuous supply of demineralised
water
• Throughput: max. permeate capacity
190 l/h
Reverse osmosis system in stainless-
steel plinth with door and floor tray
Installation of 2 x 5 l supply containers in
plinth
• 2 status LEDs (conductivity/throughflow)
• Max. yield approx. 50%
Salt filtration rate 96–98%
Water quality approx. 5–100 µS/cm
(dependent on raw water supply, typically
5–20 µS/cm)
Water connection to RO 3⁄4"
Permeate outlet 3⁄4"
Concentrate outlet 8 mm water hose
Water supply pressure 2–6 bar
Electrical connection 230 V/50 Hz
Rated load 1 kW, fuse rating 10 A
indicators
Electricity consumption: 0.6 kW/h*
• Cold water max. 28°C
max. raw water hardness 30°dGH,
15°dKH (carbonate hardness)
• Rehingeable door
• External dimensions: H 520, W 600,
D 560 mm
Accessories for reprocessing
with fully demineralised water
Single-source supplier
Water quality plays a key role in instrument
reprocessing. Raw water contains salts
and minerals which can result in deposits
on the load and on machine surfaces. Fully
demineralised water is instrumental in pre-
venting instrument corrosion. In installa-
tions requiring large volumes of water, re-
verse osmosis installations offer an eco-
nomical alternative to demineralisation
cartridges (cf. Chart on Page 41). The use
of demineralised water can increase the
economic efficiency of washer-disinfectors.
Filtration protects against deposits likely to
damage the machine, prevents downtimes
and repairs and reduces detergent costs.
The Miele range now includes the
RO-190 M1 and RO-190 M2 reverse
osmosis systems from VEOLIA.
Illustration shows washer-disinfector
with RO-190 M2 unit
Status:
Ready or Stand by
Conductivity and
throughflow
indicators:
Fault