ResMed 100 Ventilation Hood User Manual


 
13Starting therapy
English
Turning off the power
1 Stop the therapy.
2 Press the power switch at the back of the device once and follow the instructions on
the display.
Note: To disconnect the device from the mains power pull out the power cord from the
power socket.
Working with alarms
WARNING
This device is not intended to be used for vital signs monitoring. If vital signs
monitoring is required, a dedicated device should be used for this purpose.
This device is fitted with alarms to alert you to changes that will affect your treatment.
Alarm messages are displayed along the top of the screen. High priority alarms are
displayed in red, medium priority alarms in yellow and low priority alarms in light blue. The
Alarm LED lights red during high priority alarms and yellow during medium and low priority
alarms.
The alarm volume can be set Low, Medium or High. From the Setup menu, select Alarm
Settings. After the set value has been confirmed, the alarm will sound and the alarm LED
lights.
Alarm settings, see “Setup menu: Alarm Settings (Alarm Volume)” on page 16.
You can mute an alarm by pressing once. By pressing the Alarm mute key again, the
alarm sounds again. When an alarm is muted, the Alarm mute key LED will light constantly.
For a high or medium priority alarm, if after two minutes the problem is still present, the
alarm will sound again. Any active low priority alarm will be permanently muted and the
Internal Battery Use alarm will be cleared until the alarm condition is fulfilled again.
Alarm message
Alarm LED
Alarm mute key
248705 Stellar 100 and 150 User Guide EUR1.book Page 13 Monday, October 21, 2013 10:40 AM