Sanyo MPR-513 Refrigerator User Manual


 
19
ALARMS & SAFETY FUNCTIONS
Remote alarm terminal
WARNING
Always disconnect the power supply to the unit prior to connection of alarm equipment to the
terminal in order to prevent electric shock or injury.
The terminal to which a lead wire of remote alarm is connected is positioned in the box at the rear of the
cabinet. The contact capacity is DC30V, 2A.
Contact output: at normal condition; “open”
at abnormal condition; “close”
Note:
The alarm is silenced by the alarm buzzer stop key (BZ) since the remote alarm is operated in
conjunction with alarm buzzer.
Connection of remote alarm terminal
1. Remove the terminal cover at rear of cabinet.
2. Connect the lead wire of alarm equipment to the
terminal.
3. Replace the terminal cover.
terminal
terminal
cover