“Time” selected
Enter the time the test should run. The test will stop when this time is
reached.
Display will show “Min XX”. This sets the minutes to run a test. The range is
04 to 99 minutes.
“Stable” selected
Enter the conditions to be used to determine if the results are stable. The
amount of weight change to be used to determine if the test is finished,
followed by the time to be used for the weight to change. For example 0.006g
in 15 seconds would be a typical setting. Longer times or (smaller weight
changes) could be used to ensure better stability or shorter times (larger
weight changes) to ensure a faster determination. For example 0.006g in 5
seconds will end a test quicker (but possibly less accurately ) than 0.002g in
15 seconds. The user sets the weight first, then the time,
Display shows “ STAB RANGE” “ XX”. Set the range from 1 to 10 display
divisions. For the PMB 50 1 division = 0.002g, For the PMB 200 1 division =
0.02g.
Then the display shows “ Stop Time” “ Sec XX”. Range is 01 to 99 seconds.
Time/Stable will combine these 2 settings.
STOP TIME
Min 30
STABLE
St0p
STAB RANGE
02
STAB TIME
sec15
TIME/ STAB
St0p
STOP TIME
sec15
STOP RANGE
02
STOP TIME
min 30