3.3 LEVELLING THE ANALYZER
After placing the analyzer in a suitable location,
level it by using the spirit level on the rear of the
analyzer. To level the analyzer turn the two
adjustable feet at the rear of the analyzer until
the bubble in the spirit level is centred.
3.4 WARM-UP TIME
Attach the power supply cable to the connector
on the rear of the analyzer. The master power
switch is on the rear panel, next to the mains
cable. Turn the power on.
The display will indicate the analyzer serial
number and the software revision number
followed by the capacity of the analyzer. Next
the analyzer will run a self-test by displaying all
segments. The display will show zero weight
accompanied by the Î0Í symbol.
Before you start weighing, you have to wait for the analyzer to achieve a
stable internal temperature. In addition it may be necessary for some
material test to preheat the weighing chamber so that the delay to reach
operating temperature is minimised.
A stable sign
~ is shown when the analyzer is in
stable condition. It will turn off if the analyzer is
not stable.
Exact zero is shown when the Î0Í symbol
is on to the left of the display area.