The meaning of the text in the file is given below. Different test settings will create a file
with different information. Not all settings are printed if they are not used by the test
program.
Text message Description Range of Values
Result=1;
Calculation of results 0=%M, 1=%S, 2=%Atro_M,
3=%Atro_S
Heat=1;
Heating profile to be used 0=Single, 1=Step, 2=Ramp
Single Temp=100;
Temperature for single heat
setting
50C to 160C
Step temp1=050;
Temperature for first heat of a
step profile
50C to 160C, 00C is off
Step Time1=30:00;
Time for the first heat of a step
profile
Time 02:00 to 99:00
Step temp2=080;
Temperature for 2
nd
heat of a
step profile
50C to 160C, 00C is off
Step Time2=30:00;
Time for the 2
nd
heat of a step
profile
Time 02:00 to 99:00
Step temp3=050;
Temperature for 3
rd
heat of a step
profile
50C to 160C, 00C is off
Step Time3=30:00;
Time for the 3rd heat of a step
profile
Time 02:00 to 99:00
Ramp Temp=100;
Final temperature of a ramp
profile
50C to 160C
Ramp Time=30:00;
Time to reach final temperature Time 02:00 to 99:00
Rapid=0;
Rapid heat function setting, 0= Off, 1= On, 2= User
Interval Time=20;
Time period used to print interim
test results
5 sec to 99 sec
Stop mode=0;
Stop mode for drying test 0=Stable, 1=Time, 2=Stab and
Time default=2
Stop time=10;
Time to stop test time = 3 to 99 min default=20;
Stable Range=0.002;
Weight the analyzers uses to
determine if the results are stable
at the end of test.
Set from 0.001 to 0.099 for PMB
53 or 0.01 to 0.99 for PMB 202
Stable Time=10;
Time used to compare weight
during a test to determine if the
results are stable at the end of
test.
5 sec to 99 sec
Start mode=0;
Start test manually or when the
cover is closed
0=Manual 1=Auto
Name=ABCD1234
Name assigned by the user if the
test results are to be stored
Up to 10 characters of alpha-
numeric characters, plus . / -
Rapid=1;
Enable Rapid heating function 0= Off, 1= On, 2= Manual
savepos=01;
Memory location the results are
stored in.
01 to 49