Operating Information
Instructions for Use
e instructions below describe how to use the Bair Paws
Model 875 warming unit with a Bair Paws gown. For more
information about the Bair Paws warming gowns, refer to the
Instructions for Use that is shipped with the warming gowns.
1. Instruct your patient to put the warming gown on so
the opening is in the back. e gown has ties on the
side and back of the neck. e gown should be tied
loosely to help the air circulate in the gown insert.
2. Connect the warming unit hose to the warming gown
by performing the following steps.
a. Squeeze and hold the two buttons
on the sides of the hose nozzle.
b. Insert the nozzle into the properly sized hose
port in the warming gown. Refer to the
Instructions for Use that is shipped with the
warming gowns for hose port location.
c. Release the buttons to lock the hose nozzle in place.
3. Connect the warming unit to a proper power source.
4. Turn the knob on the temperature controller
clockwise to turn the warming unit ON. When you
turn the warming unit on, the heater and blower
activate. Also, the “Power On” indicator light on the
warming unit illuminates.
5. Allow the patient to adjust the temperature to a
comfortable level. e temperature and airow are
adjusted by turning the knob on the temperature
controller. e temperature adjustment varies from
approximately ambient to 43±3°C. e maximum
temperature is 46°C before the Over-temp indicator
light illuminates and audible alarm sounds.
43
º
C