Check the 3M
TM
Bair Paws
TM
system website to ensure you have the most recent version of this document. www.bairpaws.com reorder #202391A
5. Allow the warming unit to run for at least 5 minutes.
• If the “Over-temp” light illuminates again, return the warming unit to 3M Patient
Warming for service.
• Otherwise, reconnect the warming unit hose to the Bair Paws gown to resume
patient warming.
General Maintenance
Calibrating the Operating Temperatures
Calibrate the operating temperatures of the Model 875 warming unit (1) after performing
a service procedure and (2) routinely every six months. Perform the calibration in a similar
environment temperature to where the Model 875 will be used. You will need a Bair Paws
temperature test kit Model 90055 to perform the calibration. Calibration instructions are
included with the test kit.
Replacing the Filter
Replace the lter every six months or sooner if necessary. To order a replacement lter,
contact 3M Patient Warming customer service using either of the telephone numbers listed
on page 17. Instructions for replacing the lter will be included with the replacement
lter.
Cleaning the Warming Unit, Handheld Controller, Hose, and Accessories
NOTICE:
To avoid device damage:
• Do not immerse any part of the Bair Paws warming unit in liquid.
• Do not use a dripping wet cloth to clean the warming unit.
• Do not use harsh solvents to clean the warming unit.
M
1. Disconnect the warming unit from the power source before cleaning.
2. Wipe the cabinet, the temperature controller, and the outside of the hose with a damp,
soft cloth and a mild cleaning solution or antimicrobial spray.
3. Dry with a separate soft cloth.
Storage
Store all componets in a cool, dry place when not in use.