Using the Holder for the Handheld Temperature Controller
Store the handheld temperature controller in the controller holder to help prevent cord clutter
around the patient and accidental damage to the warming unit.
1. Press the holder onto the warming unit hose so it is within the patient’s reach.
WARNING:
To reduce the risks associated with thermal energy and entanglement:
• Do not leave pediatric patients unattended while warming with a Bair Paws 800 series
warming system.
• Do not allow pediatric patients to handle or adjust the temperature controller without
supervision.
2. Slide the temperature controller into the holder. It should t securely with the dial
facing the patient.
43
º
C
Temperature
controller
Controller
holder
Warming unit
hose
What to Do if the “Over-temp” Indicator Light
lluminates and Audible Alarm Sounds
If the “Over-temp” indicator light illuminates, it means that the warming unit is delivering air at
a higher temperature than the unit is calibrated to deliver. When an over-temperature condition
occurs, the color indicator light illuminates, an audible alarm will sound, and the heater will
automatically turn o (the blower will continue to run). You should perform the following steps:
1. Discontinue patient warming.
• Turn the warming unit OFF.
• Disconnect the warming unit hose from the Bair Paws warming gown.
2. Unplug the warming unit from the power source and wait 5 minutes.
3. Reconnect the warming unit to a proper power source.
4. Turn the warming unit ON and select the maximum airow and temperature setting
by turning the controller knob completely clockwise.