Chapter 3 101
Agilent E4411B Specifications and Characteristics
Inputs and Outputs
Agilent E4411B Specifications and
Characteristics
Inputs and Outputs
Internal
Front Panel
Specifications Supplemental Information
Amptd Ref
a
Amplitude reference
Frequency 50 MHz
Frequency Accuracy
Frequency reference error
b
50 Amplitude
25 dBm
c
, nominal
75 Amplitude (Option 1DP)
+28.75 dBmV
c
, nominal
a. Turn the amplitude reference signal on/off by pressing the keys: Input/Output,
Amptd Ref.
b. Frequency reference error = (aging rate × period of time since adjustment + settability +
temperature stability).
c. The internal amplitude reference actual power is stored internally.
Specifications Supplemental Information
INPUT 50
Connector Type-N female
Impedance 50 , nominal
INPUT 75 (Option 1DP)
Connector BNC female
Impedance 75 , nominal
Specifications Supplemental Information
RF OUT 50 , (Option 1DN)
Connector Type-N female
Impedance 50 , nominal
RF OUT 75 , (Option 1DQ)
Connector BNC female
Impedance 75 , nominal