82 Chapter 3
Agilent E4411B Specifications and Characteristics
Amplitude
Agilent E4411B Specifications and
Characteristics
Specifications Supplemental Information
1 dB Gain Compression
Total power at input mixer
a,b
50 MHz to 1.5 GHz
50 Ω 0 dBm
75 Ω (Option 1DP) +46.75 dBmV
a. Mixer power level (dBm) = input power (dBm) – input attenuation (dB).
b. For resolution bandwidths 1 kHz to 30 kHz, the maximum input signal amplitude must
be ≤ reference level +10 dB. (Option 1DP: For resolution bandwidths 1 kHz to 30 kHz, the
maximum input signal amplitude must be ≤ reference level +5 dB.)
Specifications Supplemental Information
Displayed Average Noise
Level
(Input terminated, 0 dB
attenuation, sample detector,
Reference Level = –70 dBm)
(75 Ω: Reference Level =
–21.24 dBmV)
50 Ω 1 kHz RBW
30 Hz VBW
100 Hz RBW
1 Hz VBW
(Option 1DR)
1 kHz RBW
30 Hz VBW
(typical)
100 Hz RBW
1 Hz VBW
(Option 1DR)
(typical)
400 kHz to 10 MHz ≤ −115 dBm ≤ −123 dBm ≤ −119 dBm ≤ −129 dBm
10 MHz to 500 MHz ≤ −119 dBm ≤ −127 dBm ≤ −121 dBm ≤ −131 dBm
500 MHz to 1.0 GHz ≤ −117 dBm ≤ −125 dBm ≤ −121 dBm ≤ −130 dBm
1.0 GHz to 1.5 GHz ≤ −113 dBm ≤ −121 dBm ≤ −118 dBm ≤ −128 dBm
75 Ω, (Option 1DP)
1 MHz to 10 MHz ≤ −63 dBmV ≤ −71 dBmV ≤ −69 dBmV ≤ −79 dBmV
10 MHz to 500 MHz ≤ −65 dBmV ≤ −73 dBmV ≤ −70 dBmV ≤ −80 dBmV
500 MHz to 1.0 GHz ≤ −60 dBmV ≤ −68 dBmV ≤ −66 dBmV ≤ −75 dBmV
1.0 GHz to 1.5 GHz ≤ −53 dBmV ≤ −61 dBmV ≤ −62 dBmV ≤ −71 dBmV