Agilent Technologies E4408B Water Dispenser User Manual


 
Chapter 3 87
Agilent E4411B Specifications and Characteristics
Amplitude
Agilent E4411B Specifications and
Characteristics
Specifications Supplemental Information
Reference Level
Range 149.9 dBm to maximum
mixer level + attenuator
setting
Resolution
Log Scale ±0.1 dB
Linear Scale ±0.12% of Reference Level
50 , Accuracy
(at a fixed frequency, a fixed
attenuator, and referenced to
35 dBm)
Reference Level (dBm) input
attenuator setting (dB)
10 dBm to > 60 dBm ±0.3 dB
60 dBm to > 85 dBm ±0.5 dB
85 dBm to 90 dBm ±0.7 dB
75 (Option 1DP), Accuracy
(at a fixed frequency, a fixed
attenuator, and referenced to
18.75 dBmV)
Reference Level (dBmV)
input attenuator setting (dB)
38.75 dBmV to
> 11.25 dBmV
±0.3 dB
11.25 dBmV to
> 26.25 dBmV
±0.5 dB
26.25 dBmV to
41.25 dBmV
±0.7 dB
Specifications Supplemental Information
Display Scale Switching
Uncertainty
Switching between
Linear and Log
±0.15 dB at reference level
Log Scale Switching No error