Agilent Technologies E4408B Water Dispenser User Manual


 
Chapter 3 85
Agilent E4411B Specifications and Characteristics
Amplitude
Agilent E4411B Specifications and
Characteristics
Specifications Supplemental Information
Absolute Amplitude
Accuracy
At reference settings
a
±0.4 dB
Overall Amplitude Accuracy
b
20 to 30 °C ± (0.6 dB + Absolute
Frequency Response)
a. Settings are: reference level 25 dBm; (75 reference level +28.75 dBmV); input attenu-
ation 10 dB; center frequency 50 MHz; RBW 1 kHz; VBW 1 kHz; amplitude scale linear
or log; span 2 kHz; sweep time coupled, signal at reference level.
b. For reference level 0 to 50 dBm; input attenuation 10 dB; RBW 1 kHz; VBW 1 kHz;
amplitude scale log, log range 0 to 50 dB from reference level; sweep time coupled; sig-
nal input 0 to 50 dBm; span 20 kHz.
Specifications Supplemental Information
RF Input VSWR (at tuned
frequency)
Attenuator setting
50
0 to 5 dB attenuation 1.55:1, characteristic
10 to 60 dB attenuation 1.35:1, characteristic
75
1 MHz to 1.1 GHz
0 to 5 dB attenuation 1.55:1, characteristic
10 to 60 dB attenuation 1.35:1, characteristic
1.1 GHz to 1.5 GHz
0 to 60 dB attenuation 2.0:1, characteristic
Input protection is tripped Open input, characteristic
Amptd Ref is On Open input, characteristic
Auto Align All is selected Open input momentarily
during retrace, characteristic