Chapter 1 27
Agilent E4403B Specifications and Characteristics
Options
Agilent E4403B Specifications and
Characteristics
Specifications Supplemental Information
Output Attenuator
Accuracy
0 dB ±0.5 dB, characteristic
8 dB ±0.5 dB, characteristic
16 dB Reference
24 dB ±0.5 dB, characteristic
32 dB ±0.6 dB, characteristic
40 dB ±0.8 dB, characteristic
48 dB ±1.0 dB, characteristic
56 dB ±1.1 dB, characteristic
Tracking Generator Output Accuracy
Relative Accuracy (Referred to –20 dBm) =
Output Attenuator Accuracy + Vernier Accuracy + Output Flatness
Absolute Accuracy =
Relative Accuracy (Referred to –20 dBm) + Absolute Accuracy at 50 MHz