Sanyo MDF-C8V Freezer User Manual


 
16
B) Function mode
Setting range : 00~33
Display range : 00~39
00, 04, 08, 13, 15, 16, 18, 19, 20, 26, 27, 31 and 33 are unavailable.
Setting procedure : In PV display, keep pressing key over 5 seconds to enter function
mode (F00 is displayed). Change the blinking 1
st
digit to desired function
code with key and key. Press SET key to be function code
available.
If you press SET key after input digits among 00, 04, 08, 13, 15, 16, 18,
19, 20, 31 and 33, the display automatically reverts to PV.
Out of setting range : If you try to input F34~F39, it is unacceptable to change with
automatically revert to PV display.
5. Warning function
High temp. alarm : When PV is higher than SV+SV
H
(high temp. alarm SV) +1, ALARM lamp
and LED display blinks, intermittent buzzer beeps with approx. 15
minutes of delay, and remote alarm output turns on.
When PV is lower than SV+ SV
H
, ALARM lamp and LED display go off,
buzzer stops beeping, and remote alarm output turns off.
If you press BUZZER key, the buzzer stops beeping instead remote
alarm output does not turn off.
Low temp. alarm : When PV is lower than SV-SV
L
(low temp. alarm SV) -1, ALARM lamp
and LED display blinks, intermittent buzzer beeps with approx. 15
minutes of delay, and remote alarm output turns on.
When PV is lower than SV- SV
L
, ALARM lamp and LED display go off,
buzzer stops beeping, and remote alarm output turns off.
If you press BUZZER key, the buzzer stops beeping instead remote
alarm output does not turn off.
6. Other function
Auto Return : If there is not any key operation for 90 seconds in each setting mode,
Key Lock mode and Function mode, the display automatically reverts to
PV. Note) Auto Return is not worked in F09 and F10.
Sensor temperature : F12: Temperature in temp. sensor is displayed
(Ex. -80.2㷄㩷 㸢㩷 displayed as ‘80.2’)
F14: Temperature of comp. sensor is displayed.
(Ex. 67㷄㩷 㸢㩷 displayed as ‘067’)
Battery accumulating
time :
F03: Battery accumulating time is displayed.
(Ex. 2years and 6months consumed 㸢㩷 displayed as ‘02.5’)
The display shows ’02.8’ to inform the battery exhaustion.
Then replace with new battery.
<Reset of accumulating time>
Step into F06. Input ‘409’ and press SET key to turn the display
to ’00.0’ with BATTERY lamp goes off.
ROM version :
F30: ROM version is displayed (Ex. Ver. 1.00 㸢㩷 displayed as “1.00”)
Condensing fan
motor accumulating
time :
F32: Notice for condensing fan motor accumulating time
000: Not notify 001: Notify
When you set in ‘001’ and fan motor accumulating time is
reached to 5.6 years, ‘FAn’ and PV are alternately displayed.
<Reset of accumulating time>
Step into F06. Input 410’ and press SET key to clear
non-volatile memory in ’00.0’. The display reverts to PV only.