4.3 Viewing data (during ventilation)
Figure 14: Viewing data
The following figure shows the screens for each option:
Figure 15: Information screens (examples)
For any further information, please contact your HME provider.
1
2
27/08/2008 17:20
cm
H2O
9
60
50
40
30
20
10
0
PS.Vs
Stop
cm
H2O
12
60
50
40
30
20
10
0
Back
PS.Vs
Menu
View settings/alarms
Technical information
Events/Alarms Log
View Monitoring
With the main screen displayed, press
to access the “Menu” screen.
Use to select the option you wish to
display:
the ventilation settings and alarm
thresholds;
the measurements (monitoring);
the technical information or;
the Event and Alarm Log, then press
to confirm.
cmH2O
10
60
50
40
30
20
10
0
cmH2O
0
60
50
40
30
20
10
0
27/08/2008 16:49
cm
H2O
0
60
50
40
30
20
10
0
PS.Vs
View alarm thresholds
min max
f -- 20
bpm
Vt 0.25 0.80
L
FiO2 21 100 %
cm
H2O
0
60
50
40
30
20
10
0
PS.Vs
PEEP 4
fmin 15
Rise t
Vts 0.50 Timax 2.0
Timin 0.3
TgE AUTO
TgI(V) NO
TgI(P) AUTO
PS 10
cmH2O
cmH2O
bpm
Ls
s
2
View settings
: Alarm thresholds
: Settings
27/06/2011 16:48
cm
H2O
0
60
50
40
30
20
10
0
DATE & TIME ALL EVENTS CODE
Events #1 to 8 of 17
: Alarms only: Scroll
/
21
25
21
01
3
2
1
0
High FiO2:%FiO2:
High FiO2:%FiO2:
Low FiO2:%FiO2:
Start of ventilation: prog:
Loss of power supply:
Switched to internal battery:
Switched to external supply:
Switched to mains supply:
28/06/2011 18:38:48
28/06/2011 18:54:40
27/06/2011 09:52:43
25/06/2011 08:13:03
24/06/2011 11:06:22
23/06/2011 09:52:23
22/06/2011 10:32:03
21/06/2011 19:55:54
27/06/2011 16:49
cm
H2O
0
60
50
40
30
20
10
0
DATE & TIME ALARM EVENTS CODE
Events #1 to 8 of 11
: All events: Scroll
/
21
25
21
--
--
--
--
--
High FiO2:%FiO2:
High FiO2:%FiO2:
Low FiO2:%FiO2:
Mains supply
Mains supply
Patient disconnection
Patient disconnection
External battery
28/06/2011 18:38:48
28/06/2011 18:54:40
27/06/2011 09:52:43
20/06/2011 08:13:03
20/06/2011 08:13:20
13/06/2011 09:52:23
13/06/2011 09:52:30
01/06/2011 19:55:54
PS.Vs
View Monitoring
f = 15 bpm FiO2 = 21 %
I:E = 1:2.3 Leaks = 0 %
Ti = 1.00 s
Vti = 0.60 L
MV
I
= 9.0 L/min
Patient hours= 1650h
Machine hours= 2000h
Turbine wear= 0%
Circuit type= Valve
Software version= x.xx
Ventilation settings and alarm thresholds:
Measurements:
Technical information:
Event and Alarm Log: