FIGURE 4.2- ANALOG OUTPUT JUMPERS
5.3 A/D - D/A Calibration Procedure
This procedure should be performed whenever a circuit board assembly is exchanged or
whenever the analog output voltage range is changed.
To calibrate the ADC, do the following:
1. Press SETUP-DIAG.
2. Enter Diagnostic password and press NEXT until D/A CALIBRATION
appears in the display and press ENTR.
3. Press ADC to perform the A/D Cal.
4. The M450H display will read "ADJUST ZERO:A/D=xx.x MV.” Put the
probe of a voltmeter between TP3(AGND) and TP9(DAC #0) on the top of
the V/F card.
5. The value displayed by the voltmeter should be close(+/- 20 mV) to the value
on the M450H display. If they are not close then the V/F card has probably
been configured improperly.
6. Adjust the Zero pot(R27) on the V/F card until the value on the M450H
display matches the value on the voltmeter to within +/- 2 mV. Note that
when adjusting R27, the value on the M450H display will change, the value
on the voltmeter will remain constant.
7. Press ENTR.
8. The M450H display will now read "ADJUST GAIN:A/D=xx.x MV.”
9. Adjust the Span Pot (R31) on the V/F card until the value on the M450H
display matches the value on the voltmeter to within +/- 2 mV.
10. Press ENTR.
11. The ADC is now calibrated and the M450H will automatically calibrate all the
DAC's. This process takes only a few seconds.
12. Press EXIT 3 times to return to the sample menu.