Agilent E4403B Specifications and Characteristics
Frequency
Agilent E4403B Specifications and
Characteristics
Specifications Supplemental Information
Frequency Span
Range 0 Hz (zero span), 100 Hz to
3GHz
Resolution 2 Hz
Accuracy ±1.0% of span
Specifications Supplemental Information
Sweep Time
Range
4 ms to 4000 s
a
Tracking Generator On
(Option 1DN)
50 ms is the minimum sweep
time
Accuracy
(Span = 0 Hz)
4 ms to 4000 s
a
±1%
Sweep Trigger
b
Free Run, Single, Line, Video
c
,
External, Delayed, Offset
d
Delayed Trigger,
e
Range 1 µs to 400 s
Resolution
rounded up to nearest µs
Accuracy ±(500 νσ + (0.01% of delay))
Offset Trigger
d
Resolution
Range ±327 ms to ±323 ks Where ST = sweep time
a. For firmware revisions prior to A.04.00, 5 ms to 2000 s.
b. Auto align is suspended in video, external, and delayed trigger modes while waiting for a
trigger event to occur.
c. Unavailable when RBW 300 Hz (Option 1DR).
d. For firmware revision A.04.00 or later.
e. Delayed trigger is available with line and external trigger.
delay in seconds
65000
----- ----------------------------------
sweep time
400
-------- -------------------
32766 ST×
400
-------- -----------------------
to
32365 ST×
400
------ ----------------------