Agilent Technologies E4403B Water Dispenser User Manual


 
56 Chapter 2
Agilent E4408B Specifications and Characteristics
Amplitude
Agilent E4408B Specifications and
Characteristics
Specifications Supplemental Information
Spurious Responses
Second Harmonic Distortion
Input Signal
10 MHz to 500 MHz < 60 dBc for 30 dBm signal
at input mixer
a
+30 dBm SHI (second
harmonic intercept)
500 MHz to 1.5 GHz < 70 dBc for 30 dBm signal
at input mixer
a
+40 dBm SHI
1.5 GHz to 2.0 GHz < 80 dBc for 10 dBm signal
at input mixer
a
+70 dBm SHI
2.0 GHz to 3.35 GHz
< 95 dBc
b
for 10 dBm signal
at input mixer
a
+85 dBm SHI
3.35 GHz to 6.6 GHz
< 95 dBc
b
for 10 dBm signal
at input mixer
a
+85 dBm SHI
6.6 GHz to 13.25 GHz
< 95 dBc
b
for 10 dBm signal
at input mixer
a
+85 dBm SHI
Third Order Intermodulation
Distortion
10 MHz to 100 MHz +5 dBm TOI (third order
intercept), characteristic
100 MHz to 3 GHz < 75 dBc for two 30 dBm
signals at input mixer
a
and
>50 kHz separation
+7.5 dBm TOI
3.0 GHz to 6.7 GHz < 75 dBc for two 30 dBm
signals at input mixer
a
and
>50 kHz separation
+7.5 dBm TOI
6.7 GHz to 13.2 GHz < 70 dBc for two 30 dBm
signals at input mixer
a
and
>50 kHz separation
+5.0 dBm TOI
13.2 GHz to 26.5 GHz < 70 dBc for two 30 dBm
signals at input mixer
a
and
>50 kHz separation
+5.0 dBm TOI
Other Input Related Spurious
Inband Responses
>30 kHz offset < 65 dBc for 20 dBm signal
at input mixer
a