Agilent Technologies E4403B Water Dispenser User Manual


 
54 Chapter 2
Agilent E4408B Specifications and Characteristics
Amplitude
Agilent E4408B Specifications and
Characteristics
Specifications Supplemental Information
Resolution Bandwidth
Switching Uncertainty
(at Reference Level)
1 kHz RBW Reference
3 kHz to 3 MHz RBW ±0.3 dB
5 MHz RBW ±0.6 dB
100 Hz to 300 Hz RBW
(Option 1DR)
±0.3 dB
Specifications Supplemental Information
Reference Level
Range 149.9 dBm to maximum
mixer level + attenuator
setting
Resolution
Log Scale ±0.1 dB
Linear Scale ±0.12% of Reference Level
Accuracy (at a fixed frequency,
a fixed attenuator, and
referenced to 30 dBm)
Reference Level (dBm) input
attenuator setting (dB)
10 dBm to > 60 dBm ±0.3 dB
60 dBm to > 85 dBm ±0.5 dB
85 dBm to 90 dBm ±0.7 dB
Specifications Supplemental Information
Display Scale Switching
Uncertainty
Switching between
Linear and Log
±0.15 dB at reference level
Log Scale Switching No error