Agilent Technologies E4403B Water Dispenser User Manual


 
16 Chapter 1
Agilent E4403B Specifications and Characteristics
Amplitude
Agilent E4403B Specifications and
Characteristics
Specifications Supplemental Information
Displayed Average Noise
Level
(Input terminated, 0 dB
attenuation, sample detector,
Reference Level = 70 dBm)
1 kHz RBW
30 Hz VBW
100 Hz RBW
1 Hz VBW
(Option 1DR)
1 kHz RBW
30 Hz VBW
(typical)
100 Hz RBW
1 Hz VBW
(Option 1DR)
(typical)
1 MHz to 10 MHz 117 dBm 126 dBm
10 MHz to 1.0 GHz 117 dBm 125 dBm 120 dBm 130 dBm
1.0 GHz to 2.0 GHz 116 dBm 124 dBm 120 dBm 130 dBm
2.0 GHz to 3.0 GHz 114 dBm 122 dBm 120 dBm 130 dBm
Specifications Supplemental Information
Display Range
Log Scale Ten divisions displayed;
0.1, 0.2, 0.5 dB/division and 1
to 20 dB/division in 1 dB steps
RBW 1 kHz Calibrated 0 to 85 dB from
Reference Level
RBW 300 Hz (Option 1DR)
Calibrated 0 to 120 dB
a
from
Reference Level
Linear Scale Ten divisions
Scale Units dBm, dBmV, dBµV, dBµA, A,
V, an d W
a. 0 to 70 dB range when span = 0 Hz, or when IF Gain fixed:
(:DISPlay:WINDow:TRACe:Y[:SCALe]:LOG:RANGe:AUTO OFF).