Agilent Technologies E4403B Water Dispenser User Manual


 
Chapter 2 49
Agilent E4408B Specifications and Characteristics
Amplitude
Agilent E4408B Specifications and
Characteristics
Specifications Supplemental Information
Displayed Average Noise
Level
(Input terminated, 0 dB
attenuation, sample detector,
Reference Level = 70 dBm)
1 kHz RBW,
30 Hz VBW
100 Hz RBW
1 Hz VBW
(Option 1DR)
1 kHz RBW
30 Hz VBW
(typical)
100 Hz RBW
1 Hz VBW
(Option 1DR)
(typical)
1 MHz to 10 MHz 117 dBm 127 dBm
10 MHz to 1.0 GHz 116 dBm 124 dBm 119 dBm 129 dBm
1.0 GHz to 2.0 GHz 115 dBm 123 dBm 120 dBm 130 dBm
2.0 GHz to 3.0 GHz 112 dBm 120 dBm 118 dBm 128 dBm
3.0 GHz to 6.0 GHz 112 dBm 120 dBm 118 dBm 128 dBm
6.0 GHz to 12 GHz 110 dBm 118 dBm 117 dBm 127 dBm
12 GHz to 22 GHz 107 dBm 115 dBm 114 dBm 124 dBm
22 GHz to 26.5 GHz 101 dBm 109 dBm 112 dBm 122 dBm
Specifications Supplemental Information
Display Range
Log Scale Ten divisions displayed;
0.1, 0.2, 0.5 dB/division and 1
to 20 dB/division in 1 dB steps
RBW 1 kHz Calibrated 0 to 85 dB from
Reference Level
RBW 300 Hz (Option 1DR)
Calibrated 0 to 120 dB
a
from
Reference Level
Linear Scale Ten divisions
Scale Units dBm, dBmV, dBµV, dBµA, A,
V, and W
a. 0 to 70 dB range when span = 0 Hz, or when IF Gain fixed:
(:DISPlay:WINDow:TRACe:Y[:SCALe]:LOG:RANGe:AUTO OFF).