Chapter 3 87
Agilent E4411B Specifications and Characteristics
Amplitude
Agilent E4411B Specifications and
Characteristics
Specifications Supplemental Information
Reference Level
Range –149.9 dBm to maximum
mixer level + attenuator
setting
Resolution
Log Scale ±0.1 dB
Linear Scale ±0.12% of Reference Level
50 Ω, Accuracy
(at a fixed frequency, a fixed
attenuator, and referenced to
−35 dBm)
Reference Level (dBm) – input
attenuator setting (dB)
–10 dBm to > –60 dBm ±0.3 dB
–60 dBm to > –85 dBm ±0.5 dB
–85 dBm to –90 dBm ±0.7 dB
75 Ω (Option 1DP), Accuracy
(at a fixed frequency, a fixed
attenuator, and referenced to
18.75 dBmV)
Reference Level (dBmV) –
input attenuator setting (dB)
38.75 dBmV to
> –11.25 dBmV
±0.3 dB
–11.25 dBmV to
> –26.25 dBmV
±0.5 dB
–26.25 dBmV to
–41.25 dBmV
±0.7 dB
Specifications Supplemental Information
Display Scale Switching
Uncertainty
Switching between
Linear and Log
±0.15 dB at reference level
Log Scale Switching No error