18 Chapter 1
Agilent E4403B Specifications and Characteristics
Amplitude
Agilent E4403B Specifications and
Characteristics
Attenuation Accuracy Relative to the 10 dB Attenuator Setting, Characteristic
Frequency Range
Attenuation dc–3.0 GHz
0 dB ±0.3 dB
5 dB ±0.3 dB
10 dB Reference
15 dB ±0.4 dB
20 dB ±0.4 dB
25 dB ±0.5 dB
30 dB ±0.5 dB
35 dB ±0.6 dB
40 dB ±0.6 dB
45 dB ±0.7 dB
50 dB ±0.7 dB
55 dB ±0.9 dB
60 dB ±0.9 dB
65 dB ±1.0 dB
Specifications Supplemental Information
Absolute Amplitude
Accuracy
At reference settings
a
±0.4 dB
Overall Amplitude
Accuracy
b
20 to 30 °C ± (0.6 dB + Absolute
Frequency Response)
a. Settings are: reference level –20 dBm; input attenuation 10 dB; center frequency 50
MHz; RBW 1 kHz; VBW 1 kHz; amplitude scale linear or log; span 2 kHz; sweep time
coupled, signal at reference level.
b. For reference level 0 to –50 dBm; input attenuation 10 dB; RBW 1 kHz; VBW 1 kHz;
amplitude scale log, log range 0 to –50 dB from reference level; sweep time coupled; sig-
nal input 0 to –50 dBm; span ≤20 κΗζ.