84 Chapter 3
Agilent E4411B Specifications and Characteristics
Amplitude
Agilent E4411B Specifications and
Characteristics
Specifications Supplemental Information
Frequency Response
a
50 Ω
9 kHz to 1.5 GHz
10 dB attenuation
20 to 30 °C ±0.5 dB
0 to 55 °C ±1.0 dB
0 dB, 5 dB, 15 to 60 dB
attenuation
±1.0 dB, characteristic
75 Ω (Option 1DP)
1 MHz to 1.5 GHz
10 dB attenuation
20 to 30 °C ±0.5 dB
0 to 55 °C ±1.0 dB
0, 5, 15 to 50 dB attenuation ±1.0 dB, characteristic
55 to 60 dB attenuation
1 MHz to 1 GHz ±1.0 dB, characteristic
1 GHz to 1.5 GHz ±1.25 dB, characteristic
a. Frequency response values are referenced to the amplitude at 50 MHz.
Specifications Supplemental Information
Input Attenuation
Switching Uncertainty at
50 MHz
Attenuator Setting
0 dB to 5 dB ±0.3 dB
10 dB Reference
15 dB ±0.3 dB
20 to 60 dB attenuation ±(0.1 dB + 0.01 × Attenuator
Setting)