Specifications
15
A = intrinsic error
En = variations
R = routine test
T = type test
[]
100% x
valuefiducial
B
%B ±=
Measuring
Range
Display Range Resolution
Instrinsic
Error
Max. Op-
erating
Error
0.001 ...2.999 0.001
3.00 ...29.99 0.01
0.020 30.0 ...299.9 0.1
300 k 0.300 k...2.999 k 1
± ( 2 % of
mv +2 digit )
± ( 5% of
mv +5
digit )
3.00 k...29.99 k 10
30.0 k...299.9 k 100
Measuring time: typ. 8 sec. with a fixed frequency
30 sec. max. with AFC and complete cycle of all
measuring frequencies
Additional error because of
probe-and auxiliary earth
electrode resistance:
digits
R
RR
E
SH
5%1025.1
)2000(
6
+××
+
Measuring error of RH and
RS:
typ. 10 % of
HsE
RRR ++
Max. probe resistance: 1 M
Max. auxiliary earth elec-
trode resistance:
1 M
Automatic check if error is kept within the limits required by IEC61557-5.
If after a measurement of probe-, auxiliary earth electrode- and earthing resis-
tance, a measurement error of higher than 30 % is assumed because of the in-