Fluke 1625 Kitchen Utensil User Manual


 
Earth/Ground Tester
Specifications
19
Resistance Measurement (R
F
)
Measuring method: current- voltage measurement as per IEC61557-4
possible
Open circuit voltage: 20 V DC
Short circuit current: 200 mA DC
Formation of measured
value:
with 4-pole measurement wires on H, S, ES can
be extended without additional error.
Resistances >1 in wire E can cause additional
error of 5m /.
Measuring
Range
Display Range Resolution Intrinsic
Error
Operating
Error
0.001 ... 2.999 0.001
0.020 3.0 ... 29.99 0.01
3 k 30.0 ... 299.9 0.1
±( 2 % of
m.v. +2
digit)
±( 5% of
m.v. +5
digit)
300 ... 2999 1
Measuring sequence: approx. 2 measurements/s
Measuring time: typ. 4 sec. incl. reversal of polarity (2 pole or 4
pole)
Max. interference voltage: 3 V AC or DC, with higher voltages measure-
ment will not be started
Max inductivity: 2 Henry
Max. overload: Urms = 250 V