Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
Remove the cuvet.
Add carefully the content of
one packet of HI 93755-0
Alkalinity Indicator Reagent.
Replace the cap and shake
vigorously for 30 seconds.
Note: Pay attention not to
spill reagent otherwise full
color development may be
inhibithed.
Replace the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press TIMER and the display will show the countdown prior to
measurement. Alternatively, wait for 2 minutes and press READ
DIRECT. In both cases “SIP” will blink during measurement.
The instrument directly displays the alkalinity in mg/L of CaCO
3
on the Liquid Crystal Display.
Alkalinity