Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
Remove the cuvet.
Add the content of one packet
of HI 93716-0 Bromine
Reagent. Replace the cap and
shake gently for about 20
seconds to dissolve most of
the reagent.
Replace the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press TIMER and the display will show the countdown prior to
measurement. Alternatively, wait for 2 minutes and 30 seconds
and press READ DIRECT. In both cases “SIP” will blink during
measurement.
The instrument directly displays concentration in mg/L of bromine
on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by: Chlorine, Chlorine Dioxide, Ozone.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake
the sample for approximately 2 minutes after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
, the
color of the sample may develop only partially, or may rapidly fade.
To resolve this, neutralize the sample with diluted HCl.
Bromine