32
10 mL
SPECIFICATIONS
Range 0.00 to 5.00 mg/L
Resolution 0.01 mg/L
Precision ±0.03 mg/L @ 1.00 mg/L
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter
@ 575 nm
Method Adaptation of the USEPA approved method. The
reaction between free copper and the bicinchoninate
reagent causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity/test
HI 93702-0 Bicinchoninate 1 packet
REAGENT SETS
HI 93702-01 Reagents for 100 tests
HI 93702-03 Reagents for 300 tests
For other accessories see page 49.
MEASUREMENT PROCEDURE
Select the program number corresponding to Free Copper on the
secondary LCD by pressing PROGRAM and .
Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace
the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
FREE COPPER
Free Copper