Remove the cuvet.
Add carefully the content of
one packet of HI 93756A-0
Ca Hardness 1 Reagent.
Replace the cap and shake
vigorously for 10 seconds.
Note: Pay attention not to
spill reagent otherwise full
color development may be
inhibithed.
Add the content of one packet
of HI 93756B-0 Ca Hardness
2 Reagent. Replace the cap
and shake vigorously for 10
seconds.
Replace the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press TIMER and the display will show the countdown prior to
measurement. Alternatively, wait for 2 minutes and press READ
DIRECT. In both cases “SIP” will blink during measurement.
The instrument directly displays concentration in mg/L of calcium
as CaCO
3
on the Liquid Crystal Display.
To convert the result to mg/L of calcium (Ca
2+
), multiply by a
factor of 0.4.
INTERFERENCES
Interference may be caused by:
Orthophosphate: precipitate Calcium at the pH of the test.
Alkalinity above 150 mg/L CaCO
3
may cause cloudiness. To resolve
this, neutralize the sample with diluted HCl.