10 mL
SPECIFICATIONS
Range 0 to 200 mg/L
Resolution 1 mg/L from 0 to 100 mg/L;
10 mg/L above 100 mg/L
Precision ±5 mg/L @ 60 mg/L
Typical EMC ±1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter
@ 525 nm
Method Turbidimetric method. The reaction between
cyanuric acid and the reagent causes a white
turbidity in the sample.
REQUIRED REAGENTS
Code Description Quantity/test
HI 93722-0 Cyanuric Acid Reagent 1 packet
REAGENT SETS
HI 93722-01 Reagents for 100 tests
HI 93722-03 Reagents for 300 tests
For other accessories see page 49.
MEASUREMENT PROCEDURE
Select the program number corresponding to Cyanuric Acid on the
secondary LCD by pressing PROGRAM and .
Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace
the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
CYANURIC ACID
Cyanuric Acid