36
10 mL
SPECIFICATIONS
Range 0 to 200 mg/L
Resolution 1 mg/L from 0 to 100 mg/L;
10 mg/L above 100 mg/L
Precision ±5 mg/L @ 60 mg/L
Typical EMC ±1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter
@ 525 nm
Method Turbidimetric method. The reaction between
cyanuric acid and the reagent causes a white
turbidity in the sample.
REQUIRED REAGENTS
Code Description Quantity/test
HI 93722-0 Cyanuric Acid Reagent 1 packet
REAGENT SETS
HI 93722-01 Reagents for 100 tests
HI 93722-03 Reagents for 300 tests
For other accessories see page 49.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Cyanuric Acid on the
secondary LCD by pressing PROGRAM ▼ and ▲.
• Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace
the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
CYANURIC ACID
Cyanuric Acid