10 mL
SPECIFICATIONS
Range 6.5 to 8.5 pH
Resolution 0.1 pH
Precision ±0.1 pH
Typical EMC ±0.1 pH
Deviation
Light Source Tungsten lamp with narrow band interference filter
@ 525 nm
Method Adaptation of the Phenol Red method. The reaction
between the sample and the reagent causes a
yellow to a red tint in the sample.
REQUIRED REAGENTS
Code Description Quantity/test
HI 93710-0 Phenol Red Indicator 1 packet
REAGENT SETS
HI 93710-01 Reagents for 100 tests
HI 93710-03 Reagents for 300 tests
For other accessories see page 49.
MEASUREMENT PROCEDURE
Select the program number corresponding to pH on the secondary
LCD by pressing PROGRAM and .
Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace
the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
pH
pH