10 mL
SPECIFICATIONS
Range 0.00 to 5.00 mg/L
Resolution 0.01 mg/L from 0.00 to 2.50 mg/L;
0.10 mg/L above 2.50 mg/L
Precision ±0.03 mg/L @ 1.00 mg/L
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter
@ 525 nm
Method Adaptation of the EPA method 330.5 and Standard
Methods for the Examination of Water and
Wastewater, 20
th
edition, 4500-Cl G. The reaction
between chlorine and the DPD reagent causes a
pink tint in the sample.
REQUIRED REAGENTS
Code Description Quantity/test
HI 93711-0 DPD Powder Reagent 1 packet
REAGENT SETS
HI 93711-01 Reagents for 100 tests
HI 93711-03 Reagents for 300 tests
For other accessories see page 49.
MEASUREMENT PROCEDURE
Select the program number corresponding to Total Chlorine on the
secondary LCD by pressing PROGRAM and .
Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace
the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
TOTAL CHLORINE
Total Chlorine