Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
Remove the cuvet.
Add the content of one packet
of HI 93702-0 Copper
Reagent. Replace the cap and
shake gently for 15 seconds.
Add the content of one packet
of HI 93702T-0 Copper Total
Reagent. Replace the cap and
shake vigorously for 15
seconds.
Replace the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press TIMER and the display will show the countdown prior to
measurement. Alternatively, wait for 45 seconds and press READ
DIRECT. In both cases “SIP” will blink during measurement.
The instrument directly displays concentration in mg/L of total
copper on the Liquid Crystal Display.