Cornelius Cold Beverage Dispenser Beverage Dispenser User Manual


 
IntelliCarb Training Manual
© 2004, IMI Cornelius Inc. - 37 - Publication Number: TP01071
Pump Troubleshooting
DOES NOT OPERATE / GAS APPLIED / DISPENSER VALVE OPEN
B-I-B empty or inlet tubing pinched off activating vacuum "sold-out".
Gas regulator over-pressurizing. (Pump stalled)
Outlet tube kinked or restricted.
Operated without fluid for excessive period. (Dry run)
Transfer tube and gas lines contaminated (syrup, rust, oil, etc.) [ensure clean gas supply, change
out all contaminated pumps]
Internal damage of control cover.
OPERATES BUT WILL NOT PRIME / DISPENSER VALVE OPEN
[consult Start-up Procedure for proper priming]
Pump valves have no moisture/dry (add water/syrup to the inlet port with pump stroking slowly).
Vacuum leaks at Q.D., barb fitting clamps, or inlet fitting o-ring.
Debris in valve seats or warped/swollen valves.
DOES NOT ACHIEVE SOLD-OUT WITH EMPTY B-I-B
Vacuum leaks at Q.D., barb fitting clamps, or inlet fitting o-ring.
Excessive amount of air in B-I-B from improper packaging.
Air trapped in outlet tubing and/or pump fluid chambers.
AIR IN INLET AND/OR OUTLET TUBING
Vacuum leaks at Q.D. o-ring or barb fitting clamps.
Vacuum leaks at inlet fitting; o-ring pinched or missing.
Large amounts of air noticed only in the outlet tubing when pump operates (diaphragm/piston assem-
blies ruptured).
STROKES WITH DISPENSER VALVE CLOSED
Air trapped in outlet tubing and/or pump fluid chambers (open outlet and purge air, check for vacuum
leaks, or air in B-I-B).
Debris in outlet valves or warped/swollen valves.
FLUID FROM EXHAUST OR VISIBLE WITHIN GAS INLET TUBING
Carbonator check valve.
Ensure clean gas supply.
Diaphragm/piston assemblies ruptured.
[change out all contaminated pumps]
GAS BLOWING FROM EXHAUST CONTINUOUSLY
Control cover subjected to contaminated gas supply or damaged (ensure clean gas supply, change
out all contaminated pumps).