ResMed 2011-09 Ventilation Hood User Manual


 
6
1 Connect the power cord.
2 Plug the free end of the power cord into a power outlet.
3 Connect the antibacterial filter firmly onto the air outlet of the device.
4
Connect the external humidifier to the other side of the antibacterial filter.
*If not using an external humidifier, the HMEF can be connected to the leak port (at step 9).
5 Connect the air tubing to the external humidifier.
6 Connect the leak port to the air tubing.
7 Select the mask type Trac h (select Setup menu, then Clinical Settings, then Advanced
Settings).
8 Perform Learn Circuit (see “Setup menu: Options” on page 16).
9 If an external humidifier is not being used, you can connect the HMEF to the leak port if
necessary.
10 Connect the catheter mount.
** The leak port or the HMEF can be
connected to standardised tracheostomy interfaces
including connector pieces like catheter mounts.
The catheter mount and the external humidifier are not part of
the ResMed component.
Antibacterial filter
External humidifier*
Air tubing
ResMed leak port
Catheter mount**
1
2
3
4
5
HMEF*
6
10
AC locking clip
AC power cord
Leak Circuit
configuration