ResMed 2011-09 Ventilation Hood User Manual


 
23Troubleshooting
English
LCD: Circuit disconnected!
The air tubing is not connected properly
to the humidifier or the device.
1
. Check that the air tubing is connected
properly to the humidifier or the
device.
2. If the alarm is n
ot cleared
, stop
treatment.
3. Re-start treatment.
LCD: Low Minute Ventilation!
Minute ventilation level has dropped
belo
w
the alarm setting level.
Contact your clinician.
LCD: Low Respiratory Rate!, High Respiratory Rate!
The respiratory rate level has dropped
below
or has exceeded the alarm setting
level.
Contact your clinician.
LCD: High Leak!
High mask leak for more than 20
seconds.
Adjust the mask to minimise leak (see
“Using mask-fit” on page 14).
Check the air circuit integrity and
reconnect.
If the p
roblem persists c
ontact your
clinician.
LCD: Non-Vented Mask!
Connection of a non-vented mask.
Mask vents may be blocked.
ResMed leak port is missing or vent
is
blocked.
Ensure the mask has vents.
Ensure the mask vents are not
block
ed.
Ensure the leak port is installed and
that the ve
nt is not blocked.
Ensure oxygen (if in use) has only been
connected at the rear of the de
vice.
If the problem persists c
ontact your
clinician.
LCD: Apnoea!
The device detects an apnoea that has
ex
ceeded the pre-set alarm level.
Breathe normally to disable the alarm.
If the problem persists c
ontact your
clinician.
LCD: Internal Battery Low!
The internal battery capacity is below
30%.
Connect the device to mains power.
LCD: Low SpO
2
!
SpO
2
has dropped below pre-set alarm
level.
Check the attachment of the sensor.
If the problem persists c
ontact your
clinician.
LCD: SpO
2
finger sensor failure!
The finger sensor is not connected
properly or delivers f
aulty values.
Check if the finger sensor is attached
properly to the finger and connection to
the pulse oximeter.
LCD: Xpod oximeter disconnected!
The pulse oximeter is disconnected.
Check if the pulse oximeter is connected
properly to the device.
Problem / possible cause Action