Transmit Selected Units U<CR> Torr<CR>
Transmit Software Version# V<CR> Hastings Instruments - Model 2002
Version 1.60 - (07-02-2002)<CR>
PP
PP
P
arameter Modificaarameter Modifica
arameter Modificaarameter Modifica
arameter Modifica
tion Commandstion Commands
tion Commandstion Commands
tion Commands
Command Description Format Valid Range:
Modify High Setpoint H={m.dd}E{+e}<CR> 1.00000e-9 to 9.99999e+9
Modify Low Setpoint L={m.dd}E{-e}<CR> 1.00000e-9 to 9.99999e+9
Modify Selected Gas # G={d}<CR> 0 To 4 (Decimal)
Modify Selected Units U={u}<CR> T, or M, {or P, if Available}
Modify Decimation Ratio D={dddd}<CR> 63 To 7936 (Decimal)
Modify RS-485 Address *{aa}A={aa}<CR> 1 To FF (Hexadecimal)
Modify Turnaround Delay *{aa}T={dd}<CR> 0 To 255 (Decimal)
Notes:Notes:
Notes:Notes:
Notes:
The setpoints may also be entered as a decimal number, e.g. [H=760.99<CR>] will be same as
entering [H=7.6099E+2<CR>] .
When inputting setpoint data, it should be entered in the same Units of Pressure as the presently
selected Units of Measurement (i.e.Torr, mbar or Pascal). The data is only checked to be a valid
number with a one digit exponent before being accepted. There are no limit checks on the data,
the user is free to choose any value appropriate to his use of the instrument.
The Turnaround Delay and RS-485 address are unique to multipoint communications. In order
to prevent inadvertent modifications of these parameters, the multipoint attention character and
the Model 2002’s present address [*{aa}] MUST be used and are checked for validity before the
command is executed.
If the RS-485 address is unknown, the ‘UNIVERSAL ADDRESS’ [*00] may be used to set the
address to a known value, e.g. [*00A=35<CR>] will change the RS-485 address to 35.
Caution:Caution:
Caution:Caution:
Caution: Since all units will respond to the ‘UNIVERSAL ADDRESS’, make sure that only
the unit to be modified is connected to the RS-485 Bus. If more than one unit is connected, this
will result in all of the units being set to the same address.
The value entered for the Turnaround Delay is used to modify an internal timer which normally
runs at ~8 millisecond, e.g. [*{aa}T=10<CR>] will set the delay to ~80 milliseconds.
If the command syntax is not met or if the number is out or range, the HPM-2002 will respond
with the ASCII codes for <bell>?<CR>, and the command will be ignored.
CalibraCalibra
CalibraCalibra
Calibra
tion tion
tion tion
tion
Adjustment CommandsAdjustment Commands
Adjustment CommandsAdjustment Commands
Adjustment Commands
Command Description Format Valid Range: