Teledyne 6712SR Refrigerator User Manual


 
6712SR Refrigerated Sampler
Section 6 SDI-12 Sondes
6-7
Use the arrow keys to select the desired parameter to
validate. Any parameter can be validated. You will be prompted
to place the probe in your validation standard. When stable,
press . The reading will be displayed at the bottom of the
screen.
When you are done validating, select “DONE” to return to the
Hardware options screens.
6.7 Disconnecting Sondes When you have disconnected all sondes, and have interrogated
the sampler for all its data, you can indicate that there is no
SDI-12 sonde.
From the main menu, select Other Functions, Hardware.
Advance through the screens until you see one that displays
“SDI-12 DATA.” Select this screen to display the following:
Select “NO SDI-12 SONDE.” You will advised that you have a
new hardware setup:
If you have collected all the data, select “DONE.” A partition
memory message will be displayed briefly, and you will be
returned to the Hardware options screens.
If you do not have a sonde attached, you may want to turn off
automatic scanning for sondes. This will eliminate the time
taken to scan for sondes whenever the sampler is turned on.
NO SDI-12 SONDE
SETUP KNOWN SONDES
SCAN SDI-12 SONDES
NEW HARDWARE SETUP--
DOWNLOAD DATA NOW
OR LOSE ALL DATA!
DONE