Section 4 Standard Programming
4-19
4.15 Reports The 6712 records a variety of data while running a sampling
program. It uses the data to produce four reports:
The Program Settings report, listing the program
settings.
The Sampling Results report, listing the program
settings, time of samples and other program events.
The Combined Results report, combining the sample
event times with rainfall or module data.
The Summary report, listing daily summaries of data
collected. The Summary report may be rainfall and/or
module data.
The sampler stores the data for each report in memory where
they remain until you select RUN. Selecting RUN clears the
memory so that it can store the data from the next program.
4.15.1 Collecting Reports There are a number of ways to collect the reports:
View reports (program settings, sampling results, and abbre-
viated summary reports) on the sampler’s display.
Collect the reports (except for summary reports) with a teledyne
Isco 581 RTD and use a computer running flowlink or samplink
to transfer the reports from the RTD to a file on the computer.
Collect the reports (except for summary reports) with an IBM PC
or compatible computer running flowlink.
Collect the reports with and IBM PC or compatible computer
running samplink.
4.15.2 Viewing the Data Because the display area is small, the data on the reports will be
displayed a little differently from the data that is transferred to
and viewed on your computer.
For the Sampling Report, the sampler displays each
program event, one at a time.
For the Module and Rainfall Data reports, it displays
daily summaries, instead of the full reports available
with the rtd, Flowlink, or Samplink.
If you select Sampling Report or Rainfall (step 3 in Viewing Data,
shown in the margin), the sampler begins displaying the report
data. The sampler advances automatically through the report
items, displaying each item briefly. While the sampler advances
automatically through the displays:
Stop the automatic displays by pressing Stop once.
Then, use the arrow keys to move manually through the
report.
Return to the main menu by pressing Stop twice.
At the end of the report, the sampler leaves the last item dis-
played until you press:
the arrow keys to move forward or backward manually
through the report.
Stop to return to the main menu.
Interrogator Icon