Teledyne 6712SR Refrigerator User Manual


 
6712SR Refrigerated Sampler
Section 5 Extended Programming
5-11
The above example presented several screens that are not
available in standard programming. Descriptions of features
available in extended programming follow.
61
RESET RAIN HISTORY?
YES NO
Select YES.
62
“LEVEL” CONDITION:
SET POINT
RANGE
RATE OF CHANGE
Select SET POINT.
63
“LEVEL” SET POINT
0.25 ft
(0.100-30.000)
Enter 0.25.
64
ONCE ENABLED,
STAY ENABLED?
YES NO
Select YES.
65
SAMPLE AT ENABLE?
YES NO
Select NO.
66
START OF SAMPLING
AFTER A
15 MINUTE DELAY
(1-999)
Enter 20.
67
PAUSE RESUME
1. HH:MM DD HH:MM DD
2. HH:MM DD HH:MM DD
CLEAR DONE
Select DONE.
Screen will say “End of Part B.”
68
NO DELAY TO START
DELAYED START
CLOCK TIME
WAIT FOR PHONE CALL
Select DELAYED START.
69
START “STORM”
AFTER A 72 HOUR
“DRY” PERIOD?
YES NO
Select NO. (Option for samplers using a 0.01 inch tip rain gauge. See Dry
Period Option, described on page 5-26.)
70
START “STORM”
AFTER A
1 MINUTE DELAY
(1-999)
Enter 1.
71
PROGRAMMING COMPLETE
RUN THIS PROGRAM
NOW?
YES NO
Select NO.
Table 5-2 Storm Water Runoff Sampling (Continued)