Teledyne 6712SR Refrigerator User Manual


 
6712SR Refrigerated Sampler
Section 4 Standard Programming
4-3
The following table is an example of flow-paced sampling.
Program type: Standard
Site description: SITE 29
Bottle kit: 24, 1000 milliliter bottles
Suction line length: 7 feet
Pacing: Flow pacing, every two pulses
Distribution: 2 bottles per sample
Sample Volume: 250 milliliters
Start time: Clock Time, 6:00 a.m. on Monday,
Wednesday, Friday
Module: No module installed
11
SAMPLE VOLUME:
200 ml (10-1000)
Type the volume of the sample you want deposited in each bottle. Then,
press
.
12
NO DELAY TO START
DELAYED START
CLOCK TIME
WAIT FOR PHONE CALL
For this example, select DELAYED START by pressing an arrow until the
option blinks. Then, press
.
13
FIRST SAMPLE
AFTER A
5 MINUTE DELAY
(1-999)
Type the delay period you want between the time you run the program and
the time the sampler takes the first sample. Then, press
.
14
PROGRAMMING COMPLETE
RUN THIS PROGRAM
NOW?
YES NO
Run the program immediately by selecting YES. Select NO if you want to
run the program later by selecting RUN from the main menu. Press
after
making your choice. In this example, NO is selected.
15
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
Run the program by selecting RUN and pressing .
Table 4-1 Standard Program: Sample Every 15 Minutes, One Sample Per Bottle (Continued)
Table 4-2 Standard Program: Flow-Paced Sampling, Two Bottles Per Sample
1
6712 SAMPLER
STANDARD PROGRAMMING
For HELP at any
screen press ? key.
2
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
Select PROGRAM.
3
SITE DESCRIPTION:
“FACTORY051“
CHANGE?
YES NO
Select NO.