GE CTR 80 Freezer User Manual


 
High voltage is used in the operation of this equipment. Severe injury or
death may result if personnel fail to observe the safety precautions. Before
working inside the equipment, turn off the power and disconnect the
power cord.
BATH FLUIDS
Fluids used in this unit may produce noxious or toxic fumes under certain
circumstances. Consult the fluid manufacturer’s MSDS (Material Safety
Data Sheet). Proper ventilation and safety precautions must be observed.
The instrument is equipped with a soft cutout (user settable firmware) and
a hard cutout (set at the factory). Check the flash point, boiling point, or
other fluid characteristic applicable to the circumstances of the unit opera
-
tion. Ensure that the soft cutout is adjusted to the fluid characteristics of
the application. As a guideline, the soft cutout should be set 10°C to 15°C
below the flash point of the bath fluid. See Section , Heat Transfer Fluid,
for specific information on bath fluids and Section , Cutout.
1.2.2
Cautions
THE DRAIN VALVE MUST BE INSTALLED ON THE BACK OF THE
BATH BEFORE ATTEMPTING TO FILL THE TANK WITH FLUID.
See Section 5.3, page 17 for drain installation instructions.
Always operate this instrument at room temperature between 41°F and
122°F (5°C to 50°C). Allow sufficient air circulation by leaving at least 6
inches (15 cm) of clearance around the instrument.
DO NOT overfill the bath. Overflowing liquid may damage the electrical
system. Be sure to allow for thermal expansion of the fluid as the bath
temperature increases. See Section 5.3, Bath Preparation and Filling, for
specific instructions.
DO NOT change the values of the bath calibration constants from the fac
-
tory set values. The correct setting of these parameters is important to the
safety and proper operation of the unit.
The refrigeration may be damaged or the lifetime shortened if the
set-point temperature is set above 60°C for more than one hour with the
refrigeration manually on. Ensure that the refrigeration is off when the
unit is used above 60°C.
The Factory Reset Sequence should be performed only by authorized
personnel if no other action is successful in correcting a malfunction. You
must have a copy of the most recent Report of Test to restore the test pa
-
rameters.
4
1 Before You Start