9.10.1 Baud Rate
The baud rate is the first parameter in the menu. The baud rate setting deter
-
mines the serial communications transmission rate.
The baud rate parameter is indicated by,
bAUd
Flashes “ ” and then displays the stetting
2400 b
Current BAUD rate
The baud rate of the serial communications may be programmed to 300, 600,
1200, 2400, 4800, or 9600 baud. Use “UP” or “DOWN” to change the baud
rate value.
4800 b
New BAUD rate
Press “SET” to set the baud rate to the new value or “EXIT” to abort the opera
-
tion and skip to the next parameter in the menu.
9.10.2 Sample Period
The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature measure-
ments transmitted from the serial interface. If the sample rate is set to 5, the in-
strument transmits the current measurement over the serial interface
approximately every five seconds. The automatic sampling is disabled with a
sample period of 0. The sample period is indicated by,
SPer
Flashes “SPEr” and then displays the setting
SP= 1
Current sample period (seconds)
Adjust the value with “UP” or “DOWN” and then use “SET” to store the sam
-
ple rate to the displayed value. “EXIT” does not store the new value.
SP= 60
New sample period
9.10.3 Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full du
-
plex or half duplex. With full duplex any commands received by the calibrator
via the serial interface are immediately echoed or transmitted back to the device
of origin. With half duplex the commands are executed but not echoed. The du
-
plex mode parameter is indicated by,
44