GE CTR 80 Freezer User Manual


 
If a High/Low voltage condition exists for longer than 5 seconds, the
compressor will be de-energized. The controller display will flash “Lo
LinE“ o and off while the condition exists.
Re-energization is automatic upon correction of the fault condition and af
-
ter a delay cycle of about 2 minutes.
High and Low Voltage Protection at 115 VAC
Voltage Cutout: +/- 12.5 % (101 - 129 VAC)
Voltage Cut-In: +/- 7.5 % (106 - 124 VAC)
High and Low Voltage Protection at 230 VAC
Voltage Cutout: +/- 12.5 % (203 - 257 VAC)
Voltage Cut-In: +/- 7.5 % (213 - 247 VAC)
See the Caution in the front of this manual for additional information.
13.2 CE Comments
13.2.1 EMC Directive
This instrument has been tested to meet the European Electromagnetic Compat-
ibility Directive ( EMC Directive, 89/336/EEC). The Declaration of Confor-
mity for your instrument lists the specific standards to which the unit was
tested.
13.2.2 Low Voltage Directive (Safety)
In order to comply with the European Low Voltage Directive (73/23/EEC), this
instrument has been designed to meet the IEC 1010-1 (EN 61010-1) and IEC
1010-2-010 (EN 61010-2-010) standards.
58
13 Trouble Shooting