9.10.4 Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa
-
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,
LF
Flashes “LF” and then displays the setting
LF= On
Current linefeed setting
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
LF= OFF
New linefeed setting
9.11 Calibration Parameters
The operator of the CTR –80 controller has access to the Hard Cutout and a
number of the bath calibration constants namely R
0
, ALPHA, DELTA, and
BETA. These values are set at the factory and should not be altered. The cor
-
rect values are important to the accuracy and proper and safe operation of the
bath. Access to these parameters is available to the user only so that in the
event that the controller memory fails the user may restore these values to the
factory settings. The user should have a list of these constants and their settings
with the manual.
Caution: DO NOT change the values of the bath calibration constants
from the factory set values. The correct setting of these parameters is im
-
portant to the safety and proper operation of the bath.
The calibration parameters menu is indicated by:
CAL Calibration parameters menu
Press “SET” five times to enter the menu.
45
9 Controller Operation