HP (Hewlett-Packard) E1368A Microwave Oven User Manual


 
Installing Switches on the HP E1369A
Because the HP E1369A Microwave Switch modules do not have
Microwave Switches installed by the factory, it is necessary to install or
connect switches before operating. Determine the number of switches, and
whether the switches are to be installed on, or connected to, the module and
proceed as instructed below.
Installing Switches
on the Module
Installation described for up to three coaxial switches (channels 00-02) is
below. Figure 2-3 shows the wiring diagram and mounting hole centers for
each switch installed.
HP SWITCHES. The following 5V HP 333XX series and HP 876X series
switches will function in the HP E1369A Microwave Switch module.
Note The corresponding HP 876X and 333XX series switches shown in the table
below are electrically and physically identical. The HP 8761, 8766 (33363),
8767 (33364),8768 (33365),8769 (33366) series coaxial switches will not
function in the HP E1369A Microwave Switch module because the switch
coils are not split (separate for each contact).
HP Part Number Frequency Characteristic
Impedance
Ports Internal 50
Termination
Auto Coil
Interrupt
MTG
Hole
33311B/011 or 8762B/005/100
33311C/011 or 8762C/005/100
33312B/011 or 8763B/005/100
33312C/011 or 8763C/005/100
33313B/011 or 8764B/005/100
33313C/011 or 8764C/005/100
33314A/011 or 8765A/005/100
33314B/011 or 8765B/005/100
33314C/011 or 8765C/005/100
DC to 18 GHz
DC to 26.5 GHz
DC to 18 GHz
DC to 26.5 GHz
DC to 18 GHz
DC to 26.5 GHz
DC to 4 GHz
DC to 20 GHz
DC to 26.5 GHz
50
50
50
50
50
50
50
50
50
3
3
4
4
5
5
3
3
3
All ports
All ports
One port
One port
No ports
No ports
No ports
No ports
No ports
Yes
Yes
Yes
Yes
Yes
Yes
No
No
No
C
C
D
D
D
D
B
B
B
NON-HP SWITCHES. K&L Microwave Inc., Dynatech Microwave
Technology Inc., and RLC Electronics Inc. currently manufacture split coil
coaxial switches that will physically mount in the HP E1369A Microwave
Switch module rear panel. All switches mount using the "A" hole centers
(see Figure 2-3). When selecting switches, refer to the diagram shown in
Figure 1-1 to verify that the switch will electrically function in the HP
E1369A Microwave Switch module. Remember, switches must have split
coil operation with 42V maximum drive level.
Caution The maximum current that the control circuit can accommodate
is 1 amp per switch. Maximum current also depends on the
output capacity of the mainframe or power supply used.
20 Configuring the HP E1368A/69A/70A Modules Chapter 2