Valco electronic Valve Oven Microwave Oven User Manual


 
Chapter 2
Installing Valves in the Valve Box Valve Installation
Operating Manual 31
Installing Valves in the Valve Box
This version of Valve Oven is used when any valves heating is required.
Valve Box for up to Six Valves
This option is analogous to the Valve Oven but without heating elements.
Hence, for the valves installation procedure, follows all the steps described
previously for the heated oven and unheated compartment.
Refer to:
Installing Valves in the Heated Oven
Installing Valves in the Unheated Compartment
Valve Box for up to Two Valves
This option does not contain any heating elements. Besides, it is designed to
accomodate only up to two valves.It does not include any electronics and the
solenoid valves controlling the valves actuators are droven by the TRACE GC
Ultra.
The solenoid valves are connected to the timed event output located on the Mother
Board of the TRACE GC Ultra.
Sampling Valve 1 must be connected to
J51
Sampling Valve 2 must be connected to J52
Then, the valves installation follows the same procedure described previously for
the heated oven and unheated compartment.
Refer to:
Installing Valves in the Heated Oven
Installing Valves in the Unheated Compartment