Emerson Process Management 7951 Ventilation Hood User Manual


 
Appendix C Technical data for the 7951
Page C.15
Earthing requirements for group 1 connections only
In general, the earthing arrangements are different for large and small installations. (A small installation may
possibly consist of just one instrument.)
If the 7951 is part of a large installation with separate earths for chassis and
instrumentation:
In this case you may (depending on the overall system requirements) earth the
7951 chassis and instrumentation separately by cutting the link on the
connector board.
If the 7951 is on its own or in a small installation with one common earth for
chassis and instrumentation:
In this case you must leave the link intact so that the chassis and
instrumentation are earthed to the same point.
Link
Processor
Board
Mother
Board
Rear
Panel
Top o
f
instrument
case
Socket
S
K1
Figure C.3: Where to find the link on the connector board